Page 92
STAT
IM
2000/2000S Service Guide
96-108639 Rev 3.0
4. STAT
IM
2000 Calibration
3. Validation thermocouple calibration
To calibrate the validation thermocouple:
t Turn the power switch to OFF. Ensure the calibration jumper is in place and turn the power
switch ON.
t To start a validation thermocouple (self-calibrating) cycle, press and hold the
UNWRAPPED cycle button on the control box and press the START button. An asterisk
* will appear next to the validation thermocouple offset value (CVO) to indicate that a
validation assembly calibration cycle is running.
t If the asterisk * does not appear, press the STOP button on the control box twice to reset
the unit. Turn the power switch OFF then ON again.
t Hold down the UNWRAPPED cycle button on the control box and press the START button
to start the self-calibration again.
t This calibration will take approximately 6 minutes.
t Allow the self-calibration to complete.
t The temperature within the chamber will rise to the sterilization temperature. Wait until
sterilization phase of the calibration cycle ends automatically.
t The offset value in the upper left-hand corner of the display (CVO) may have changed to a
new offset value.
t Press the STOP button on the control box when finished.
4. Switch off the unit and remove all calibration equipment.
STATIM 2000/2000S/2000 G4 Service Guide