Page 96
STAT
IM
2000/2000S Service Guide
96-108639 Rev 3.0
4. STAT
IM
2000 Calibration
2. Validation thermocouple calibration
To calibrate the validation thermocouple:
1. Turn the unit off OFF.
2. Turn ON the unit while keeping Unwrapped and Wrapped button pressed to enter STAT
IM
Service Mode.
3. This Service Mode is password protected, enter password to continue (default password
is: Unwrapped, Wrapped, R&P and Stop keys pressed in this order).
4. Toggle using keypad through the menu selection to reach Calibration option and press
R&P key.
5. The display should show the calibration screen as before
6. To start a validation thermocouple (self-calibrating) cycle, press and hold the
UNWRAPPED cycle button on the keypad/control box and press the START button. An
asterisk “*” will appear next to the validation thermocouple offset value (CVO) to indicate
that a Validation Assy. calibration cycle is running.
7. If the asterisk “*” does not appear, press the STOP button on the control box twice to
reset the unit. Turn the power switch OFF then ON again.
8. Hold down the UNWRAPPED cycle button on the control box and press the START button
to start the self-calibration again.
9. This calibration will take approximately 6 minutes.
10. Allow the self-calibration to complete.
11. The temperature within the chamber will rise to the sterilization temperature. Wait until
sterilization phase of the calibration cycle ends automatically.
12. The offset value in the upper left-hand corner of the display (CVO) may have changed to a
new offset value.
13. Press the STOP button on the control box when finished.
3. Switch off the unit and remove all calibration equipment.
STATIM 2000/2000S/2000 G4 Service Guide