Model 4200A-SCS Prober and External Instrument Control Section 6: Using a Model 82 C-V System
4200A-913-01 Rev. A December 2020 6-49
Flatband capacitance and flatband voltage
The Model 82 uses the flatband capacitance method of finding flatband voltage, V
FB
. The Debye
length is used to calculate the ideal value of flatband capacitance, C
FB
. Once the value of C
FB
is
known, the value of V
FB
is interpolated from the closest V
G
values (Nicollian and Brews 487-488).
The method used is invalid when interface trap density becomes very large (1012-1013 and greater).
However, this algorithm should give satisfactory results for most users. Those who are dealing with
high values of D
IT
should consult the appropriate literature for a more appropriate method.
Based on doping, the calculation of C
FB
uses N at 90% W
MAX
, or user-supplied N
A
(bulk doping for
p-type, acceptors) or N
D
(bulk doping for n-type, donors).
C
FB
is calculated as follows:
Where:
C
FB
= flatband capacitance (pF)
C
OX
= oxide capacitance (pF)
S
= permittivity of substrate material (F/cm)
A = gate area (cm
2
)
1 × 10
-4
= units conversion for
1 × 10
-12
= units conversion for C
OX
And = extrinsic Debye length =
Where:
kT = thermal energy at room temperature (4,046 × 10
-21
J)
q = electron charge (1.60219 × 10
-19
coulombs)
N
x
= N at 90% W
MAX
, or N
A
, or N
D
when input by the user
N at 90% W
MAX
is chosen to represent bulk doping

Table of Contents