Section 3: Configure and use a Series 700 Switching System Model 4200A-SCS Prober and External Instrument Control
3-14 4200A-913-01 Rev. A December 2020
4. Select Save.
5. Select the second connectpins action you added to the project tree.
6. Make the following connections using the pairs of TermIdStr# and Pin# text fields in the action:
CVH1 Pin 1
CVL1 Pin 2
CVL1 Pin 3
CVL1 Pin 4
When you are finished, the Key Parameters for the action look like the following figure.
Figure 43: Second connectpins connections
7. Select Save.