Operating Instructions Model 100E Instruction Manual
116 045150102 Rev XB1
To disable a data channel, follow the instruction shown in section 6.10.2.2 then press:
Edit Data Channel Menu
SETUP X.X
0)
CONC: ATIMER, 2, 900
PREV NEXT INS DEL
EDIT
PRNT
EXIT
FROM THE DAS PASSWORD MENU
Exits to the main
Data Acquisition
menu
SETUP X.X
NAME:CONC
<SET
SET>
EDIT
PRINT EXIT
SETUP X.X
CHANNEL ENABLE:ON
<SET SET>
EDIT
PRINT EXIT
ENTR
accepts the new
setting and returns to the
previous menu.
EXIT
ignores the new setting
and returns to the previous
menu.
SETUP X.X
CHANNEL ENABLE:ON
OFF
ENTR EXIT
Toggle key to turn
channel ON or OFF
Press
SET>
key until…
6.10.2.11. HOLDOFF Feature
The iDAS HOLDOFF feature allows to prevent data collection during calibrations and
during the DAS_HOLDOFF period enabled and specified in the VARS (Section 6.6). To
enable or disable the HOLDOFF, follow the instruction shown in section 6.10.2.2 then
press:
Edit Data Channel Menu
SETUP X.X
0)
CONC: ATIMER, 2, 900
PREV NEXT INS DEL
EDIT
PRNT
EXIT
FROM THE DAS PASSWORD MENU
Exits to the main
Data Acquisition
menu
SETUP X.X
NAME:CONC
<SET
SET>
EDIT
PRINT EXIT
SETUP X.X
CAL HOLD OFF:ON
SET>
EDIT
PRINT EXIT
ENTR
accepts the new
setting and returns to the
previous menu.
EXIT
ignores the new setting
and returns to the previous
menu.
SETUP X.X
CAL HOLD OFF:ON
ON
ENTR EXIT
Toggle key to turn
HOLDOFF ON or OFF
Press
SET>