EPA Protocol Calibration Model 100E Instruction Manual
144 045150102 Rev XB1
fully warmed up and its operation has stabilized. During the calibration, the M100E should
be in the CAL mode, and therefore sample the test atmosphere through all components
used during normal ambient sampling and through as much of the ambient air inlet
system as is practicable. If the Instrument will be used on more than one range (i.e.
DUAL or AUTO ranges), it should be calibrated separately on each applicable range.
Calibration documentation should be maintained with each analyzer and also in a central
backup file.
Table 8-1: Activity Matrix for Calibration Equipment & Supplies
Equipment/supplies Acceptance limits
Frequency and method
of measurement
Action if requirements
are not met
Recorder Compatible with
output signal of
analyzer; min. chart
width of 150 mm (6
in) is recommended
Check upon receipt Return equipment to
supplier
Sample line and
manifold
Constructed of PTFE
or glass
Check upon receipt Return equipment to
supplier
Calibration equipment Meets guidelines of
reference 1 and
Section 2.3.2 (Q.A.
Handbook)
See Section 2.3.9 (Q.A.
Handbook)
Return equipment/
supplies to supplier or
take corrective action
Working standard SO
2
cylinder gas or SO
2
permeation tube
Traceable to NIST-
SRM meets limits in
traceability protocol
for accuracy and
stability (Section
2.0.7, Q.A.
Handbook)
Analyzed against NIST-
SRM; see protocol in
Section 2.0.7, Q.A.
Handbook
Obtain new working
standard and check for
traceability
Zero air Clean dry ambient
air, free of
contaminants that
cause detectable
response with the
SO
2
analyzer.
See Section 2.9.2 (Q.A.
Handbook)
Obtain air from another
source or regenerate.
Record form Develop standard
forms
N/A Revise forms as
appropriate
Audit equipment Must not be the same
as used for calibration
System must be checked
out against known
standards
Locate problem and
correct or return to
supplier