APPENDIX A-2: Setup Variables For Serial I/O,
Revision B.3
Model 100E Instruction Manual
230 045150110 Rev B.3
APPENDIX A-2: Setup Variables For Serial I/O, Revision B.3
Table A-1: M100E Setup Variables, Revision B.3
Setup Variable Numeric
Units
Default Value Value Range Description
DAS_HOLD_OFF Minutes 15 0.5 to 20 Duration of DAS hold off
period.
TPC_ENABLE — ON,
OFF
4
OFF, ON ON enables temperature
and pressure
compensation; OFF
disables it.
RCELL_SET ºC 50
Warnings:
45 to 55
30 to 70 Sample chamber
temperature set point and
warning limits.
IZS_SET
1
ºC 50
Warnings:
45 to 55
30 to 70 IZS temperature set point
and warning limits.
DYN_ZERO OFF OFF, ON ON enables contact
closure dynamic zero; OFF
disables it.
DYN_SPAN OFF OFF, ON ON enables contact
closure dynamic span;
OFF disables it.
CONC_PRECISION 1 AUTO, 0, 1, 2,
3, & 4
Number of digits to
display to the right of the
decimal point for
concentrations on the
display. Enclose value in
double quotes (“) when
setting from the RS to 232
interface.
CLOCK_ADJ Sec./Day 0 -60 to 60 Time-of-day clock speed
adjustment.
LANGUAGE_SELECT — ENGL ENGL,
SECD,
EXTN
Selects the language to
use for the user interface.
Enclose value in double
quotes (“) when setting
from the RS-232
interface.
MAINT_TIMEOUT Hours 2 0.1 to 100 Time until automatically
switching out of software-
controlled maintenance
mode.
CONV_TIME 33 MS 33 MS,
66 MS,
133 MS,
266 MS,
533 MS,
1 SEC,
2 SEC
Conversion time for PMT
and UV detector channels.
Enclose value in double
quotes (“) when setting
from the RS-232
interface.