Instrument Maintenance Model 100E Instruction Manual
156 045150102 Rev XB1
9.1. Maintenance Schedule
Table 9-1 is the recommended maintenance schedule for the M100E. Please note that in
certain environments with high levels of dust, humidity or pollutant levels some
maintenance procedures may need to be performed more often than shown.
Table 9-1: M100E Preventive Maintenance Schedule
Item Action Frequency Cal
Check ?
Manual
Section
Date Performed
TEST
functions
Review and
evaluate
Weekly No
Appendix
C
Zero/span
check
Evaluate
offset and
slope
Weekly Yes 7.3, 7.6
Zero/span
calibration
Zero and
span
calibration
Every 3
months
Yes
7.2, 7.4,
7.5
PMT Sensor
Hardware
Calibration
Low-level
hardware
calibration
On PMT or
preamp
changes
Yes 11.6.3
Particulate
filter
Change
particle
filter
Weekly No 9.3.1
Pneumatic
sub-system
Check for
leaks in
gas flow
paths
Annually or
after repairs
involving
pneumatics
Yes on
leaks, else
no
11.5.1;
11.5.2
Sample
Chamber
windows and
Filters
Clean As necessary Yes - -
External zero
air scrubber
(optional)
Exchange
chemical
Every 3
months
No 9.3.3

Table of Contents