2.2. EPA Equivalency Designation
The Model 100E Analyzer is designated as Reference Method Number EQSA-0495-100 as
per 40 CFR Part 53 when operated under the following conditions:
• Range: Any range from 50 parts per billion (ppb) to 10 parts per million (ppm).
• Ambient temperature range of 5
o
C to 40
o
C.
• Line voltage range of 105-125 VAC or 220-240 VAC, at 50 or 60 Hz.
• Sample filter: Equipped with PTFE filter element in the internal filter assembly.
• Sample flow of 650 +/- 65 cc/min.
• Vacuum pump (internal or external) capable of 14"Hg Absolute pressure @ 1 slpm
or better.
• Software settings:
Dynamic span OFF
Dynamic zero OFF
Dilution factor OFF
AutoCal ON or OFF
Dual range ON or OFF
Auto-range ON or OFF
Temp/Pressure compensation ON
Under the designation, the analyzer may be operated with or without the following
optional equipment:
• Rack mount with chassis slides.
• Rack mount without slides, ears only.
• Zero/span valve options.
• Internal zero/span (IZS) option with:
a) SO
2
permeation tube - 0.4ppm at 0.7 liter per minute;
certified/uncertified.
b) SO
2
permeation tube - 0.8 ppm at 0.7 liter per minute;
certified/uncertified. Under the designation, the IZS option cannot be used
as the source of calibration.
• 4-20mA isolated analog outputs.
• Status outputs.
• Control inputs.
• RS-232 output.
• Ethernet output.
• Zero air scrubber.
• 4-20mA, isolated output.