Getting Started Model 100E Instruction Manual
38 045150102 Rev XB1
3.2.3. Warning Messages
Because internal temperatures and other conditions may be outside of specified limits
during the analyzer’s warm-up period, the software will suppress most warning
conditions for 60 minutes after power up.
If warning messages persist after 60 minutes, investigate their cause using the
troubleshooting guidelines in Section 11. The following table includes a brief description
of the various warning messages that may appear.
Table 3-7: Possible Warning Messages at Start-Up
MESSAGE MEANING
ANALOG CAL WARNING
The instruments A/D circuitry or one of its analog outputs is not
calibrated.
AZERO WRN XXX.X MV
The reading taken during the auto-zero cycle is outside of specified limits.
The value XXX.X indicates the auto-zero reading at the time of the
warning.
BOX TEMP WARNING
The temperature inside the M200E chassis is outside the specified limits.
CANNOT DYN SPAN
Remote span calibration failed while the dynamic span feature was set to
turned on
CANNOT DYN ZERO
Remote zero calibration failed while the dynamic zero feature was set to
turned on
CONFIG INITIALIZED
Configuration was reset to factory defaults or was erased.
DARK CAL WARNING
Dark offset above limit specified indicating that too much stray lightis
present in the sample chamber
.
DATA INITIALIZED
iDAS data storage was erased.
FRONT PANEL WARN
Firmware is unable to communicate with the front panel.
HVPS WARNING
High voltage power supply for the PMT is outside of specified limits.
IZS TEMP WARNING
On units with IZS options installed: The permeation tube temperature is
outside of specified limits.
PMT DET WARNING PMT detector output outside of operational limits.
PMT TEMP WARNING
PMT temperature is outside of specified limits.
RCELL PRESS WARN
Sample chamber pressure is outside of specified limits.
RCELL TEMP WARNING
Sample chamber temperature is outside of specified limits.
REAR BOARD NOT DET
The CPU is unable to communicate with the motherboard.
RELAY BOARD WARN
The firmware is unable to communicate with the relay board.
SAMPLE FLOW WARN
The flow rate of the sample gas is outside the specified limits.
SAMPLE PRESS WARN Sample pressure outside of operational parameters.
SYSTEM RESET
The computer was rebooted.
UV LAMP WARNING
The
UV lamp intensity measured by the reference detector reading
too low or too high