Operating Instructions Model 100E Instruction Manual
74 045150102 Rev XB1
6.7.1. Accessing the Diagnostic Features
To access the DIAG functions press the following keys:
From this point
forward,
EXIT
returns
to the
SECONDARY
SETUP MENU
EXIT
returns
to the
PRIMARY
SETUP MENU
DIAG
PRESSURE CALIBRATION
PREV
NEXT
ENTR
EXIT
DIAG
OPTIC TEST
PREV
NEXT
ENTR
EXIT
SETUP X.X
ENTER DIAG PASS: 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
EXIT
DIAG
SIGNAL I / O
NEXT
ENTR
EXIT
DIAG
ANALOG OUTPUT
PREV
NEXT
ENTR
EXIT
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
DIAG
TEST CHAN OUTPUT
PREV
ENTR
EXIT
DIAG
ELECTRICAL TEST
PREV
NEXT
ENTR
EXIT
DIAG
LAMP CALIBRATION
PREV
NEXT
ENTR
EXIT
EXIT
returns
to the main
SAMPLE
display
DIAG
FLOW CALIBRATION
PREV
NEXT
ENTR
EXIT
SAMPLE* RANGE = 500.000 PPB SO2 =X.XXX
< TST TST > CAL
SETUP
6.7.2. Signal I/O
The signal I/O diagnostic mode allows to review and change the digital and analog
input/output functions of the analyzer. See Appendix A-4 for a complete list of the
parameters available for review under this menu.
NOTE
Any changes of signal I/O settings will remain in effect only until the signal
I/O menu is exited. Exceptions are the ozone generator override and the flow
sensor calibration, which remain as entered when exiting.