EasyManua.ls Logo

Teledyne 100E - Electrical Test; Lamp Calibration

Teledyne 100E
258 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating Instructions Model 100E Instruction Manual
86 045150102 Rev XB1
6.7.6. Electrical Test
The electrical test function creates a current, which substitutes the PMT signal, and feeds
it into the preamplifier board. This signal is generated by circuitry on the pre-amplifier
board itself and tests the filtering and amplification functions of that assembly along with
the A/D converter on the motherboard. It does not test the PMT itself. The electrical test
should produce a PMT signal of about 2000 ±1000 mV.
To activate the electrical test press the following keys.
SAMPLE* RANGE = 500.000 PPB SO2 =X.XXX
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
EXIT
DIAG
SIGNAL I / O
PREV
NEXT
JUMP
ENTR
EXIT
DIAG ELEC
RANGE = 500.000 PPB O2=X.XXX
<TST TST>
EXIT
SETUP X.X ENTER DIAG PASS: 818
8 1 8
ENTR
EXIT
Press
NEXT
until…
DIAG
ELECTRICAL TEST
PREV NEXT
ENTR
EXIT
Press
TST
until…
DIAG ELEC
PMT = 1732 MV
SO2=X.XXX
<TST TST>
EXIT
While the electrical test is
activated,
PMT
should equal:
2000 mV ± 1000 mV
6.7.7. Lamp Calibration
An important factor in accurately determining SO
2
concentration is the amount of UV
light available to transform the SO
2
into SO
2
* (see Section 10.1.1). The model 100E
compensates for variations in the intensity of the available UV light by adjusting the SO
2
concentration calculation using a ratio (LAMP RATIO)that results from dividing the
current UV lamp (UV LAMP) intensity by a value stored in the CPU’s memory
(LAMP_CAL). Both LAMP Ration and UV Lamp are test functions viewable from the
instruments front panel.

Table of Contents

Related product manuals