Model 100E Instruction Manual Calibration Procedures
045150102 Rev XB1 131
STEP THREE: Perform the zero/span calibration:
M-P CAL
STABIL=X.XXX PPB
SO2 =X.XXX
< TST TST >
ENTR
CONC
EXIT
M-P CAL
STABIL=X.XXX PPB
SO2 =X.XXX
< TST TST >
CAL
SETUP
M-P CAL
STABIL=X.XXX PPB
SO2 =X.XXX
< TST TST >
ZERO
CONC EXIT
EXIT
returns to the main
SAMPLE
display
ACTION:
Allow zero gas to enter the sample port at the
rear of the instrument.
M-P CAL
STABIL=X.XXX PPB
SO2 =X.XXX
< TST TST >
SPAN
CONC EXIT
M-P CAL RANGE = 500.0 PPB SO2 =X.XXX
< TST TST >
ENTR
SPAN CONC
EXIT
M-P CAL RANGE = 500.0 PPB SO2 =X.XXX
< TST TST > ENTR CONC
EXIT
Wait until
STABIL
falls below
0.5 ppb.
This may take several
minutes.
The
SPAN
key now appears
during the transition from
zero to span.
You may see both keys.
SAMPLE* RANGE = 500.000 PPB SO2 =X.XXX
< TST
TST
> CAL
SETUP
SAMPLE
STABIL=X.XXX PPB
SO2 =X.XXX
< TST TST > CAL
SETUP
Set the Display to show the
STABIL
test function.
This function calculates the
stability of the NO/NO
x
measurement
Press
ENTR
to changes the
OFFSET
&
SLOPE
values for the
SO
2
measurements.
Press
EXIT
to leave the calibration
unchanged and return to the
previous menu.
ACTION:
Allow span gas to enter the sample port at the
rear of the instrument.
The value of
STABIL
may jump
significantly.
Wait until it falls back
below
0.5 ppb.
This may take several
minutes.
Press
ENTR to
change the
OFFSET
&
SLOPE
values for the
SO
2
measurements.
Press
EXIT
to leave the calibration
unchanged and return to the
previous menu.