Calibration Procedures Model 100E Instruction Manual
134 045150102 Rev XB1
Step Three: Perform the calibration or calibration check according to the following flow
chart:
Wait until
STABIL
falls below 0.05
ppb. This may
take several
minutes.
ZERO CAL M
STABIL=XXX.X PPB SO2 =X.XXX
< TST TST >
ZERO
CONC
EXIT
returns to the
main
SAMPLE
display
ZERO CAL M STABIL=XXX.X PPB SO2 =X.XXX
< TST TST > ENTR CONC EXIT
Pressing
ENTR
changes the calibration of the instrument.
If either the
ZERO
or
SPAN
button fails to
appear, see Chapter 11
for troubleshooting tips.
Analyzer enters
ZERO
CAL
mode
.
EXIT
returns the unit to
SAMPLE
mode without
changing the calibration
values.
ZERO CAL M STABIL=XXX.X PPB SO2 =X.XXX
< TST TST > ZERO CONC
EXIT
ZERO CAL M STABIL=XXX.X PPB SO2=X.XX X
< TST TST > CAL
CALZ
CALS
SETUP
SPAN CAL M
STABIL=XXX.X PPB SO2 =X.XXX
< TST TST >
SPAN
CONC EXIT
SPAN CAL M STABIL=XXX.X PPB SO2 =X.XXX
< TST TST >
ENTR
CONC
EXIT
EXIT
returns to the
SAMPLE
mode without
changing the calibration
values.
SPAN CAL M STABIL=XXX.X PPB SO2 =X.XXX
< TST TST >
SPAN
CONC EXIT
ACTION:
Allow zero gas to enter the sample port at the
rear of the instrument.
SAMPLE* RANGE = 500.000 PPB SO2 =X.XXX
< TST
TST >
CAL
CALZ CALS SETUP
SAMPLE
STABIL=XXX.X PPB
SO2 =X.XXX
< TST TST > CAL
CALZ
CALS SETUP
Scroll the display to the
STABIL
test function. This
function calculates the stability
of the NO/NO
X
measurements.
The value of
STABIL
may jump
significantly. Wait
until it falls below
0.5
ppb. This may take
several minutes.
Pressing
ENTR
changes the calibration of the instrument.
Analyzer enters
SPAN
CAL
Mode.