Model 100E Instruction Manual Calibration Procedures
045150102 Rev XB1 137
STEP TWO: Perform the zero/span check.
Wait until STABIL
falls below 0.5
ppb. This may
take several
minutes.
The value of STABIL
may jump
significantly. Wait
until
STABIL
falls
below
0.5 ppb. This
may take several
minutes.
SPAN CAL M STABIL=XXX.X PPB SO2 =X.XXX
< TST TST > ZERO SPAN CONC EXIT
SAMPLE STABIL=XXX.X PPB SO2 =X.XXX
< TST TST > CAL
CALZ
CALS SETUP
ZERO CAL M STABIL=XXX.X PPB SO2 =X.XXX
< TST TST > ZERO CONC EXIT
SAMPLE STABIL=XXX.X PPB SO2 =X.XXX
< TST TST > CAL CALZ CALS SETUP
EXIT
returns to the main
SAMPLE
dis
la
ACTION:
Record the
SO
2
readings
presented in the
upper right corner of
the display.
ACTION:
Record the
SO
2
readings
presented in the
upper right corner of
the display.
SAMPLE* RANGE = 500.000 PPB SO2 =X.XXX
< TST TST > CAL CALZ CALS SETUP
Scroll to the STABIL test
function.