Model 100E Instruction Manual Calibration Procedures
045150102 Rev XB1 141
To program the sample sequence shown above:
SAMPLE* RANGE = 500.000 PPB SO2 =X.XXX
< TST TST > CAL CALZ CZLS
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG
ACAL
DAS RNGE PASS CLK MORE EXIT
SETUP X.X
SEQ 1) DISABLED
NEXT
MODE EXIT
SETUP X.X
SEQ 2) DISABLED
PREV NEXT
MODE
EXIT
SETUP X.X
MODE: DISABLED
NEXT
ENTR EXIT
SETUP X.X
MODE: ZERO
PREV
NEXT
ENTR EXIT
SETUP X.X
MODE: ZERO–SPAN
PREV NEXT
ENTR
EXIT
SETUP X.X SEQ 2) ZERO–SPAN, 1:00:00
PREV NEXT MODE
SET
EXIT
SETUP X.X
TIMER ENABLE: ON
SET>
EDIT EXIT
SETUP X.X
STARTING DATE: 01–JAN–02
<SET SET>
EDIT
EXIT
SETUP X.X
STARTING DATE: 01–JAN–02
0
4 SEP
0
3
ENTR
EXIT
Toggle keys
to set
day, month &
year:
Format :
DD-MON-Y
Y
SETUP X.X
STARTING DATE: 04–SEP–03
<SET
SET>
EDIT EXIT
Default
value is
ON
SETUP C.4
STARTING DATE: 04–SEP–03
<SET
SET>
EDIT EXIT
SETUP C.4
STARTING TIME:00:00
<SET SET>
EDIT
EXIT
SETUP C.4
STARTING TIME:00:00
1 4 : 1 5
ENTR
EXIT
Toggle keys to
set time:
Format : HH:MM
This is a 24 hr
clock .
PM hours are
13 – 24.
Example
2:15 PM = 14:15
Toggle keys
to set
number of
days
between
procedures
(1-367)
Toggle keys
to set
delay time for
each iteration
of the
sequence:
HH:MM
(0 – 24:00)
SETUP C.4
DELTA DAYS: 1
<SET SET>
EDIT
EXIT
SETUP C.4
DELTA DAYS: 1
0 0 2
ENTR
EXIT
SETUP C.4
DELTA DAYS:2
<SET SET> EDIT EXIT
SETUP C.4
DELTA TIME00:00
<SET SET>
EDIT
EXIT
SETUP C.4
DELTA TIME: 00:00
0 0 :3 0 ENTR EXIT
SETUP C.4 DELTA TIEM:00:30
<SET
SET>
EDIT EXIT
SETUP C.4
STARTING TIME:14:15
<SET
SET>
EDIT EXIT
EXIT returns
to the SETUP
Menu
SETUP C.4
DURATION:15.0 MINUTES
<SET SET>
EDIT
EXIT
SETUP C.4 DURATION 15.0MINUTES
3 0 .0 ENTR EXIT
SETUP C.4
DURATION:30.0 MINUTES
<SET
SET>
EDIT EXIT
Toggle keys
to set
duration for
each
iteration of
the
sequence:
Set in
Decimal
minutes
from
0.1 – 60.0
SETUP C.4
CALIBRATE: OFF
<SET SET>
EDIT
EXIT
SETUP C.4 CALIBRATE: OFF
ON
ENTR
EXIT
SETUP C.4
CALIBRATE: ON
<SET SET> EDIT
EXIT
Toggle key
between
Off and
ON
SETUP C.4
SEQ 2) ZERO–SPAN, 2:00:30
PREV NEXT MODE SET
EXIT
Sequence Delta Time
MODE Delta Days
With dynamic calibration turned on, the state of the internal setup variables
DYN_SPAN and DYN_ZERO is set to ON and the instrument will reset the slope and
offset values for the SO
2
response each time the AutoCal program runs. This continuous
re-adjustment of calibration parameters can often mask subtle fault conditions in the
analyzer. It is recommended that, if dynamic calibration is enabled, the analyzer’s test