Model 100E Instruction Manual EPA Protocol Calibration
045150102 Rev XB1 149
conducted every two weeks. Level 2 checks should be conducted in between the Level 1
checks at a frequency desired by the user. Span concentrations for both levels should be
between 70 and 90% of the measurement range.