EasyManua.ls Logo

Abbott CELL-DYN 3000 - Page 162

Abbott CELL-DYN 3000
542 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CELL-DYN 3000 System Operator’s Manual 5-33
9140240E — May 1995
Chapter 5 Operating Instructions
The COMPUTER SETUP Screen is used to configure the transmission
format to meet the requirements of the Laboratory Information System or
on-line computer. Instructions for using this option are given after the
following description of the soft keys.
The [REINIT INTERFACE] key is used to initialize the RS-232
interface for the displayed transmission configuration after it is entered.
The interface isautomatically initialized whenever the [STOP
TRANSMISS] key is pressed.
NOTE: Refer to the Interface Specification for complete
information on interfacing or calling Customer Support Center at
1 (800) CELL DYN.
The [STOP TRANSMISS] key stops the current data transmission to the
on-line computer. When the [STOP TRANSMISS] key is pressed, the
following soft key labels are displayed:
CONFIRM STOP
CANCEL STOP
These keys confirm or cancel the Stop Transmission command.
The [TOGGLE ON/OFF] key enables or disables the first five options
in the list displayed on the COMPUTER SETUP Screen.
The numbers on the COMPUTER SETUP Screen shown in Figure5.17
correspond to the following numbered options:
1.Auto-transmission of ALERTED parameter data
When this option is enabled, a report is automatically transmitted
to the LIS for any sample with flagged parameter results.
2.Auto-transmission of NON-ALERTED parameter data
When this option is enabled, a report is automatically transmitted
to the LIS for any sample without flagged parameter results.
3.Auto-transmission of ALERTED graphic data
When this option is enabled, histograms and scatterplots are
automatically transmitted to the LIS for any sample with flagged
results.