CELL-DYN 3000 System Operator’s Manual 5-63
9140240E — May 1995
Chapter 5 Operating Instructions
Figure 5.31: Run Screen for Background Counts
The [BACKGROUND] key is used to select the run mode and display
the RUN Screen for background counts. (See Figure 5.31.) Results from
this run option are identified by the designation BACKGROUND in the
data log and are automatically excluded from the X-B analysis.
CLEAR
ORIFICE
WORK
LIST
SPECIMEN
TYPE
CUSTOMIZE
REPORT
CHANGE
SAMPLER
PRINT
TICKET
PRINT
REPORT
MAIN
RUN
Ready
Dec 18 1992
Operator ID
Sequence #
09:51
734
0630
Type BACKGROUND
Param Set: 1
WBC: K/uL
NEU: %N
LYM: %L
MONO: %M
EOS: %E
BASO: %B
RBC: M/uL
HGB: g/dL
HCT: %
MCV: fL
MCH: pg
MCHC: g/dL
RDW: %
PLT: K/uL
MPV: fL
Open Sampler
D
E
P
O
L
S
I
Z
E
COMPLEXITY ORTHOGONAL
PLT RBC
BACK-
GROUND
Search
Go Back
Book TOC