5-98 CELL-DYN 3000 System Operator’s Manual
9140240E — May 1995
Operating Instructions Chapter 5
Using the Work List with Bar Code OFF
When the Bar Code OFF option is selected, the samples must be run in
the order in which the information has been entered in the Work List,
since the Work List is accessed sequentially as the samples are
processed. The sample identification number must be entered in the
Work List <SPECIMEN ID> field. The number entered in the Work List
<SPECIMEN ID> field is used to identify the sample on the Data
Station RUN Screen and in the Data Log.
CAUTION: If a sample needs to be repeated, turn the Work
List OFF before running the sample again. When the repeat run
is completed, turn the Work List ON and continue to process
samples in the order in which they were entered in the Work List.
NOTE: QC samples should not be entered in the Work List. To
run QC samples, turn the Work List OFF and refer to the “Daily
QC Checks for the CS” section of this chapter for instructions
for running QC samples.
Sample Analysis Procedure
1.Follow the instructions in the Work List Set Up Procedure to
configure the Work List.
2.From the main WORK LIST Screen, press [WORK LIST ON] to
enable the Work List.
3.If necessary, press [BAR CODE OFF] to disable the bar code
function.
4.Type the appropriate sample identification information in the
<SPECIMEN ID> and <SPECIMEN NAME> (if selected) entry
fields on the Work List. Press the Enter key on the keyboard to save the
entry in each field and advance the cursor.
5.If selected, type the appropriate information in the <LIMIT> and
<PARAM> entry fields. Press the Enter key on the keyboard to save
the entry in each field and advance the cursor.
NOTE: If no entries are made in these fields, the default
selections are used.
6.After the information for the sample is entered, place the sample in
a rack or on a mixer. Subsequent samples should be placed in the
rack or on the mixer in the order in which the information is
entered into the Work List.
7.When all entries have been made, press [RETURN] to return to the
Data Station RUN Screen.
8.Press [SPECIMEN TYPE] followed by [PATIENT].