6-56 CELL-DYN 3000 System Operator’s Manual
9140240E — May 1995
Calibration Chapter 6
CELL-DYN 3000
Manual Calibration Worksheet
Instrument:____________________________________ Date:___________________________________
Operator:______________________________________
________________________________________________________________________________________
_
Calculate All Factors To Three Decimal Places
* Current factor as printed in this manual calibration procedure.
** If factor exceeds limits, do not calibrate. Check all calculations and call the Customer Support Center for assistance at 1 (800)
CELL DYN.
New Open Mode Calibration Factors
Assay or Ref Mean x Current Open Mode Cal Factor* = New Open Mode Cal Factor
Open Mode
Assay Value
or
Ref Mean /
Open
Mode
Mean x
Current
Open Mode
Cal Factor* =
New Open
Mode Cal
Factor Range**
WBC / x = 0.700 - 1.300
RBC / x = 0.800 - 1.200
HGB / x = 0.700 - 1.300
MCV / x = 0.700 - 1.300
PLT / x = 0.700 - 1.300