CELL-DYN 3000 System Operator’s Manual 6-57
9140240E — May 1995
Chapter 6 Calibration
* Current Open Mode factor as printed at the start of this calibration procedure.
* Delete the sign of the % Diff before entering it on the chart.
** Do not calibrate. Call the Customer Support Center for assistance at 1 (800) CELL DYN.
Factor % Difference
New Open Mode Factor - Current Open Mode Factor x 100 = % Diff
New Open Mode Factor
New Open
Mode
Factor -
Current Open
Mode Factor* /
New Open
Mode Factor x 100 = %Diff
WBC - / x 100 =
RBC - / x 100 =
HGB - / x 100 =
MCV - / x 100 =
PLT - / x 100 =
Manual Calibration Criteria
%Diff*
Validation Range
Cal Not Required
Cal Range
Cal Required
Cal Limit
Do Not Cal**
Cal?
Y/N
WBC <1.5%
>1.5% but <10%
>10%
RBC <1.0%
>1.0% but <10%
>10%
HGB <1.0%
>1.0% but <10%
>10%
MCV <1.0%
>1.0% but <10%
>10%
PLT <3.0%
>3.0% but <15%
>15%