CELL-DYN 3000 System Operator’s Manual 6-87
9140240E — May 1995
Chapter 6 Calibration
CELL-DYN 3000
Manual Calibration Worksheet
Instrument:____________________________________ Date:___________________________________
Operator:______________________________________
________________________________________________________________________________________
_
Calculate All Factors To Three Decimal Places
* Delete the sign of the % Diff before entering it on the chart.
** Do not calibrate. Call the Customer Support Center for assistance at 1 (800) CELL DYN.
Mode to Mode Calibration Difference
Closed Mode Mean - Open Mode Mean x 100 = % Diff
Open Mode Mean
Closed Mode
Mean -
Open Mode
Mean /
Open Mode
Mean x 100 = % Diff
WBC - / x 100 =
RBC - / x 100 =
HGB - / x 100 =
MCV - / x 100 =
PLT - / x 100 =
Mode to Mode Calibration Criteria
%Diff*
Validation Range
Cal Not Required
Cal Range
Cal Required
Cal Limit
Do Not Cal**
Cal?
Y/N
WBC <1.75%
>1.75% but <10%
>10%
RBC <1.25%
>1.25% but <10%
>10%
HGB <1.25%
>1.25% but <10%
>10%
MCV <1.25%
>1.25% but <10%
>10%
PLT <3.50%
>3.50% but <20%
>20%