6-88 CELL-DYN 3000 System Operator’s Manual
9140240E — May 1995
Calibration Chapter 6
* Current factor as printed in this manual calibration procedure.
** If factor exceeds limits, do not calibrate. Check all calculations and call the Customer Support Center for assistance at 1 (800)
CELL DYN.
* Mean after calibration.
** If % Diff exceeds limits, call the Customer Support Center for assistance at 1 (800) CELL DYN.
New Closed Mode Calibration Factors
Open Mode Mean x Current Closed Mode Cal Factor* = New Closed Mode Cal Factor
Closed Mode Mean
Open Mode
Mean /
Closed
Mode
Mean x
Current
Closed Mode
Cal Factor* =
New
Closed
Mode Cal
Factor Range**
WBC / x = 0.700 - 1.300
RBC / x = 0.800 - 1.200
HGB / x = 0.700 - 1.300
MCV / x = 0.700 - 1.300
PLT / x = 0.700 - 1.300
Mode to Mode Post-Calibration Difference
Closed Mode Mean - Open Mode Mean x 100 = % Diff
Open Mode Mean
Closed
Mode
Mean* -
Open
Mode
Mean /
Open Mode
Mean x 100 = %Diff Range**
WBC - / x 100 = <1.75%
RBC - / x 100 =
<1.25%
HGB - / x 100 =
<1.25%
MCV - / x 100 =
<1.25%
PLT - / x 100 =
<3.50%