10-42 CELL-DYN 3000 System Operator’s Manual
9140240E — May 1995
Troubleshooting Chapter 10
If necessary, refer to the directions given in Chapter 5, Operating
Instructions, for customizing the display.
Obtain printouts of the WBC kinetic data for several samples.
From the first DIAGNOSTICS MENU Screen, press [CNT
RATE SUMMARY].
Press [WBC CNT RATE] to display the data for the kinetic
rate followed by [PRINT] to obtain a printout.
Press [WBC RATE GRAPH] to display the graph of the
kinetic data followed by [PRINT] to obtain a printout.
If there is a flagging problem or a problem with an abnormal
sample, obtain a printout of a normal sample for comparison.
Obtain a printout of the RAW DATA SUMMARY Screen
immediately after the problem sample is run. From the first
DIAGNOSTICS MENU Screen, press [RAW DATA
SUMMARY] followed by [PRINT] to obtain a printout.
6. RBC, HGB, MCV and PLT:
Configure the RUN Screen to display the RBC and PLT
histograms. Obtain printouts of problem samples.
Obtain a printout of the RAW DATA SUMMARY Screen
immediately after the problem sample is run. From the first
DIAGNOSTICS MENU Screen, press [RAW DATA
SUMMARY] followed by [PRINT] to obtain a printout.