10-42 CELL-DYN 3000 System Operator’s Manual
9140240E — May 1995
Troubleshooting Chapter 10
If necessary, refer to the directions given in Chapter 5, Operating
Instructions, for customizing the display.
• Obtain printouts of the WBC kinetic data for several samples.
From the first DIAGNOSTICS MENU Screen, press [CNT
RATE SUMMARY].
Press [WBC CNT RATE] to display the data for the kinetic
rate followed by [PRINT] to obtain a printout.
Press [WBC RATE GRAPH] to display the graph of the
kinetic data followed by [PRINT] to obtain a printout.
• If there is a flagging problem or a problem with an abnormal
sample, obtain a printout of a normal sample for comparison.
• Obtain a printout of the RAW DATA SUMMARY Screen
immediately after the problem sample is run. From the first
DIAGNOSTICS MENU Screen, press [RAW DATA
SUMMARY] followed by [PRINT] to obtain a printout.
6. RBC, HGB, MCV and PLT:
• Configure the RUN Screen to display the RBC and PLT
histograms. Obtain printouts of problem samples.
• Obtain a printout of the RAW DATA SUMMARY Screen
immediately after the problem sample is run. From the first
DIAGNOSTICS MENU Screen, press [RAW DATA
SUMMARY] followed by [PRINT] to obtain a printout.