10-50 CELL-DYN 3000 System Operator’s Manual
9140240E — May 1995
Troubleshooting Chapter 10
Table 10.1 Instrument Status Conditions (Continued)
Bar Code flag not allowed to change
unless Work List is purged/BL
––––––––––––––Explanation/Action–––––––––––––––
The [BAR CODE ON] or [BAR CODE OFF] key was pressed and
there is an existing work list.
Delete all samples from the work list before turning the bar code ON or
OFF.
__________________________________________________________
Auto Sampler Off/BL
––––––––––––––Explanation/Action–––––––––––––––
Power to the Sample Loader is turned OFF.
Turn ON the power switch and wait for the initialization cycle to be
completed.
__________________________________________________________
Auto-Sampler Initializing/BL
––––––––––––––Explanation/Action–––––––––––––––
The Sample Loader hardware initialization cycle is in progress.
Wait for the cycle to be completed.
__________________________________________________________
Auto Sampler Ready/BL
––––––––––––––Explanation/Action–––––––––––––––
The Sample Loader initialization cycle is complete and samples may be
processed. Press the Sample Loader START key to initiate processing.
__________________________________________________________
Auto Sampler Pause/BL
––––––––––––––Explanation/Action–––––––––––––––
The Sample Loader PAUSE key was pressed during operation and
therefore, operation is suspended.
Press the Sample Loader START key to resume operation.