CELL-DYN 3000 System Operator’s Manual 10-51
9140240E — May 1995
Chapter 10 Troubleshooting
Table 10.1 Instrument Status Conditions (Continued)
Auto Sampler Busy/BL
––––––––––––Explanation/Action––––––––––––––
An action was requested during Sample Loader operation and the
Sample Loader cannot perform it.
Press the Sample Loader PAUSE key before requesting the desired
action.
__________________________________________________________
Samples Completed/BL
––––––––––––Explanation/Action––––––––––––––
The Sample Loader has completed processing samples in the end rack
and has stopped automatically.
__________________________________________________________
Change Sampler in Ready State Only/BL
––––––––––––Explanation/Action––––––––––––––
The [CHANGE SAMPLER] key was pressed while the Analyzer was
busy.
The [CHANGE SAMPLER] key can only be pressed when the
Analyzer is in the READY state.
__________________________________________________________
Change sampler when Auto Sampler is not busy/BL
––––––––––––Explanation/Action––––––––––––––
The [CHANGE SAMPLER] key was pressed while the Sampler Loader
was busy.
Press the Sample Loader PAUSE key before pressing [CHANGE
SAMPLER].
__________________________________________________________
Change Specimen Type when Auto-Sampler is not busy/BL
––––––––––––Explanation/Action––––––––––––––
The [SPECIMEN TYPE] key was pressed while the Sample Loader
was busy. Press the Sample Loader PAUSE key before pressing
[SPECIMEN TYPE].