10-72 CELL-DYN 3000 System Operator’s Manual
9140240E — May 1995
Troubleshooting Chapter 10
Table 10.3 Sample-Related Fault Conditions (Continued)
RBC Metering Fault — Clog or Flow Error/BL
CLOG or FLOW ERROR is displayed on the RUN Screen to the right
of the RDW result.
RBC CLOG or RBC FLOW is printed on the graphics report.
Results for RBC, PLT and related parameters are suppressed.
–––––Probable Cause––– ––––Corrective Action–––
1. The timing for the RBC/PLT 1. Repeat the sample. If the
metering was outside problem was caused by
acceptable limits. debris in the aperture, the
automatic cleaning cycle
may have corrected it.
2. Clean the RBC/PLT
aperture as directed in
Chapter 9, Maintenance.
NOTE: The Analyzer automatically cleans the RBC/PLT
aperture after a metering fault is detected. CLEANING RBC
ORIFICE is displayed in the status box.