6240B DC Voltage Current Source/Monitor Operation Manual
4.2.11 SHIFT/LOCAL Key (Shift Mode/Local)
4-17
4.2.11 SHIFT/LOCAL Key (Shift Mode/Local)
SHIFT key (in normal operation): Functions as SHIFT key, and the SHIFT indicator goes ON.
In the shift status, function names printed in blue characters on the
panel are enabled. Press the SHIFT key again to release the shift
status.
LOCAL key (in remote control): Releases remote control.
Turns OFF the RMT indicator and switches from remote control
to panel operation.
NOTE: Remote control is not released if the LLO (Local Lock Out) command is set through the GPIB or USB
interface.
4.2.12 STBY Key (Output Standby)
Turns off the output relays to set Standby status. The OPR indicator goes OFF.
NOTE: Whenever switching between Operate and Standby, the output relays are turned on and off every time.
To extend the relay life spans, using the Suspend function that switches between Operate and Suspend
is recommended.