6240B DC Voltage Current Source/Monitor Operation Manual
9. SPECIFICATIONS
9-1
9. SPECIFICATIONS
All accuracy specifications are guaranteed for one year at a temperature of 23 5 C and a relative humidity
not exceeding 85 %.
9.1 Source and Measurement
However, the measurement resolution with integration time of 100 s, 500 s or S/H (Sample Hold) is as
follows:
The source range in the 4 A range is limited by the duty factor condition.
For 4 A range pulse source, maximum pulse width 20 ms/duty factor 20 %
Range Source range
Setting
resolution
Measurement range
Measurement
resolution
Voltage source/
measurement range
300 mV 0 to 320.00 mV 10 V0 to 320.999 mV 1 V
3 V 0 to 3.2000 V 100 V0 to 3.20999 V 10 V
15 V 0 to 15.000 V 1 mV 0 to 15.1999 V 100 V
Current source/
measurement range
30 A0 to 32.000 A 1 nA 0 to 32.0999 A 100 pA
300 A0 to 320.00 A 10 nA 0 to 320.999 A1 nA
3 mA 0 to 3.2000 mA 100 nA 0 to 3.20999 mA 10 nA
30 mA 0 to 32.000 mA 1 A0 to 32.0999 mA 100 nA
300 mA 0 to 320.00 mA 10 A0 to 320.999 mA 1 A
1 A 0 to 1.0000 A 100 A0 to 1.01999 A 10 A
4 A 0 to 4.0000 A 200 A0 to 4.0
1999 A 10 A
Resistance
measurement
Determined by
voltage range/
current range
calculations
- - 0 to 0.75 G Minimum 
Integration time
100 s500 sS/H (100 s)
Measurement resolution (digits)
10 2 10