6240B DC Voltage Current Source/Monitor Operation Manual
6.7.4 TER? Command
6-40
6.7.4 TER? Command
The TER? command reads out the self test results.
1. Command response
2. Descriptions of values a, b, c, d and e
The TER register column in Table 5-19 shows the error factors and register values for a, b, c, d and e.
For example, an error in VSVM 3 V +FS of the self test is returned as follows.
00000, 00000, 00016, 00000, 00000
Source
Operate/Standby H Sets the output to OFF (Standby).
E Sets the output to ON (Operate).
E?, H? Returns the present output status.
Response:
Random sweep
memory data
(using D com-
mand)
N [adr] Random sweep memory data setting starts from
the N command and completes at the P command.
0
*6
P N<adr>,D<data1><UNIT>,D<data2>
<UNIT>,...,P
adr: Memory address (0 to 7999)
data1:Voltage or current source value
at adr
data2:Voltage or current source value
at adr+1
*6: Not initialized by the RINI or *RST command.
Item Command Description
Default Operation
Power
ON
Factory
shipment
DC/pulse
OPR/SUS
Sweep
OPR/SUS
Status E?, H?
Operate E
Suspend H
Standby H
Note:
1. When no source range is specified, the optimal
range is set.
2. Source values different from the present source
function cannot be set.
ddddd,ddddd,ddddd,ddddd,ddddd
a
bc
d
e