6240B DC Voltage Current Source/Monitor Operation Manual
6.7.3 Remote Command List
6-24
6.7.3 Remote Command List
1. The Default column shows default settings at Power ON or at factory shipment.
• The Power ON column shows the status when the power is turned ON.
• The *RST or RINI command initializes the settings to factory default.
However, the RINI command cannot initialize
*5
and the RINI or *RST command cannot initial-
ize
*6
.
2. Note for description in the command list
• Parameter enclosed in [ ] can be omitted.
• Parameters enclosed in < > are single delimited data items.
• The symbols in the Operation column shows the following:
: Acceptable
: Not acceptable
in DC pulse OPR/SUS column:
Acceptable only in the HOLD trigger mode or in Suspend status
in Sweep OPR/SUS column:
Acceptable only when sweep stops or in Suspend status.
: Acceptable in the Suspend status.
Item Command Description
Default Operation
Power ON
Factory
shipment
DC/pulse
OPR/SUS
Sweep
OPR/SUS
Source
Source mode MD0 DC mode
MD1 Pulse modem
MD2 DC sweep mode
MD3 Pulse sweep mode
MD4 Low-resistance measurement pulse mode
MD? Response: MD0 to MD4
Source function VF Voltage source function
Executing
this func-
tion sets
Suspend.
Executing
this func-
tion sets
Suspend.
IF Current source function
V? Response: V3 to V5 for VF
I-1 to I5 for IF
I?
Source range SVRX Optimal range
SVR3 300 mV range
SVR4 3 V range
SVR5 15 V range
SVR? Response: SVRX3 to SVRX5 (for the optimal range)
SVR3 to SVR 5 (for the fixed range)