6240B DC Voltage Current Source/Monitor Operation Manual
9.4 General Specifications
9-12
9.4 General Specifications
Operating environment conditions: Temperature 0 ºC to +50 ºC, relative humidity 85 % or below with no
condensation
Storage environment conditions: Temperature -25 ºC to +70 ºC, relative humidity 85 % or below with
no condensation
Warming up time: 60 minutes or longer (until it settles in the specified accuracies.)
Display: 16 segments ï‚´ 12 digits vacuum fluorescent display
Power supply: AC power 100 V, 120 V, 220 V, and 240 V (User selectable)
Line frequency: 50 Hz/60 Hz
Power consumption: 85 VA or less
Dimensions: Approx. 212 (width) ï‚´ 88 (height) ï‚´ 400 (depth) mm
Mass: 5 kg or less
Safety: Compliant with IEC61010-1 Ed.3
EMC: Compliant with EN61326-1 classA
Vibration proof: Compliant with IEC60068-2-6 2G
Option No. Standard OPT.32 OPT.42 OPT.44
Power voltage 100 V 120 V 220 V 240 V
Specify the option number when ordering.
Use a power cable and a fuse that are compliant with the safety
standard when changing the power supply voltage.