6240B DC Voltage Current Source/Monitor Operation Manual
5.2.3 Operations in Sweep Source Mode
5-15
5.2.3.1 Operations in DC Sweep Source Mode
The following table shows operations in the DC sweep source mode.
Table 5-5 Operations in DC Sweep Source Mode (1/2)
Operational
condition
Trigger mode Description Operation Remarks
Operate ON
AUTO
Executes continu-
ous measurement
with the setting
period time Tp.
Th: Hold time
Tp: Period time
Td: Measurement delay time
Tds:Source delay time
Tm: Measurement time
(Integration time +
Measurement data
processing time)
Tcn: Operate processing time
Trc: Range change processing
time
HOLD
Executes mea-
surement after
trigger input.
Changing the
source range
AUTO
Changes the
source range dur-
ing sweep.
HOLD
Th
Tp
Tp
Td
Tm
Td
Tm
Tds Tds
Tds
3
2
1
Bias
Operate
Start trigger
COMPLETE OUT [HI/GO/LO]
SYNC OUT
COMPLETE OUT [END]
COMPLETE OUT [FRONT]
TRIGGER IN
2
Operate
Th
Td Tm
TmTd
Tp
Tds
1
Start trigger
Bias
Trigger
SYNC OUT
COMPLETE OUT [FRONT]
TRIGGER IN
COMPLETE OUT [HI/GO/LO]
COMPLETE OUT [END]
Tp
Tds
Th
Bias
1
2
3
Tp
Tp-Tds
Trc
Td
Tm
Td-
Tm
Tds
Tds
Start trigger
Changes the source range
Operate
SYNC OUT
COMPLETE OUT [FRONT]
TRIGGER IN
COMPLETE OUT [HI/GO/LO]
COMPLETE OUT [END]
Tds
Operate
Th Tp
Tm
Trc
Tm
TdsTs
Td-Tds
1
Start trigger
Bias
Trigger
SYNC OUT
COMPLETE OUT [FRONT]
TRIGGER IN
COMPLETE OUT [HI/GO/LO]
COMPLETE OUT [END]
Td
Tds
Tp-Tds