6240B DC Voltage Current Source/Monitor Operation Manual
8. CALIBRATION
8-1
8. CALIBRATION
This chapter describes how to calibrate the 6240B to perform within the specified accuracy ranges.
In order to use the 6240B in the specified accuracies, periodic calibration at least once a year is recom-
mended.
Contact an ADC CORPORATION sales representative for the calibration service.
8.1 Measuring Instrument and Cables Necessary for Calibration
The following table shows the required accuracies of a DMM and cables used for calibration.
Range
ZERO FS
Cable
Calibration
point
Required
accuracy
Calibration
point
Required
accuracy
300 mV 0 V 1.5 V 300 mV 15 ppm A01044
(supplied accessory) *1
3 V 5 V 3 V 10 ppm
15 V 50 V 15 V 20 ppm
30 A0 A 50 pA30 A 120 ppm
300 A 500 pA 300 A 120 ppm
3 mA 5 nA 3 mA 120 ppm
30 mA 50 nA 30 mA 120 ppm
300 mA 500 nA 300 mA 210 ppm
1 A 5 A 1 A 170 ppm
4 A 5 A 1 A 170 ppm
*1: When much externally induced noise exists, use shielded cables, such as A01001, etc.