6240B DC Voltage Current Source/Monitor Operation Manual
2.2.4 Initializing Setting Conditions
2-33
2.2.4 Initializing Setting Conditions
The following procedure initializes the 6240B to the factory default settings.
However, the following items are not initialized.
• Selected interface
• GPIB address
• Talk Only/Addressable
• Header output
•USB ID
• Parameter load at power ON
Operation Character display area
1.
Press MENU, and select K) PARAMETER with
or .
K) PARAMETER
2.
Press to enter the Select level.
1) Parm Load
Load0
3.
Press to enter the Input/Run level.
1) Parm Load
Ent Load0
4.
Select "Ld Dflt" with .
1) Parm Load
Ent Ld Dflt
5. Press ENTER.
6. When loading is complete, "Done" is displayed and
the menu reverts to the Select level.
1) Parm Load
Done
7. Press EXIT to exit the menu.