6240B DC Voltage Current Source/Monitor Operation Manual
8.5.3 Voltage Measurement Calibration
8-12
11. Check the full-scale calibration value.
Fine-adjust the calibration coefficient by XUP and XDN.
12. Move to the next step
When changing the voltage calibration mode: XNXT
When moving to voltage measurement calibration:
XNXT ï‚® XVM
8.5.3 Voltage Measurement Calibration
1. Move to the voltage measurement calibration mode by XVM.
2. Select the range.
300 mV range: XR3
3 V range: XR4
15 V range: XR5
3. Enter the DMM data input mode by XDAT.
4. Set the DMM read value by XD data.
5. Read out and check the measurement data.
6. Move to the full-scale calibration mode by XNXT.
7. Set the DMM read value by XD data.
8. Read out and check the measurement data.
9. Move to the next step
When changing the voltage range: XNXT
When moving to current source or voltage limiter calibration:
XNXT ï‚® XIS, XILH, XILL