6240B DC Voltage Current Source/Monitor Operation Manual
1.5 Operating Check
1-12
6. Press OPR.
The OPR indicator goes ON and the VSVM measurement starts. (See Figure
1-10.)
Figure 1-10 VSVM 0 V Measurement in 300 mV Range
7. Verify that the VM measured value is within 150 V of 0 V in the VS 300 mV
range.
8. Press STBY.
The OPR indicator goes OFF and the 6240B enters the Standby (output OFF) sta-
tus.
The operation check is complete.