6240B DC Voltage Current Source/Monitor Operation Manual
5.2.3 Operations in Sweep Source Mode
5-14
2. Changing the sweep measurement parameters
The sweep measurement parameters are basically changeable only in Standby or Suspend status, but
the following parameters are changeable during sweep stop in Operate status.
Time parameter
Hold time
Source delay time
Measurement delay time
•Pulse width
Period time
Start address and stop address for random sweep
(These are changeable in the start address to stop address range when switching from Standby
status to Operate or Suspend.)
Sweep function parameter
Repeat count
Reverse mode ON or OFF
RTB ON or OFF
Measurement auto range ON or OFF
Measurement ON or OFF
Measurement integration time
Selecting COMPLETE OUT or SYNC OUT external control signal output function
3. Indicator display for sweep status
indicates the sweep status.
Sweeping: The indicator rotates.
HOLD: The indicator stops rotation but remains displayed.
STOP: The indicator is not displayed.