6240B DC Voltage Current Source/Monitor Operation Manual
5.2.12 Operating Multiple 6240B
5-64
Thus, consider the Tsync time and make the relevant settings as follows:
1. Tds2 Tds1 - Tsync
2. Td2 Td1 - Tsync
3. Tw2 Tw1
4. Tp2 Tp1-T
A
Measurement
trigger
Source
SYNC OUT
6240B
(Master)
SYNC OUT TRIGGER IN
6240B (Master) 6240B (Slave)
(Set to HOLD)
Measurement
Tp1
Tds1
Tw1
Td1
Tsync
Tp2
Tds2
Tw2
Td2
TRIGGER IN
Source
SYNC OUT
6240B
(Slave)
Measurement
Internal
measurement
trigger
Tsync
Measure-
ment
Measure-
ment
Measure-
ment
Measure-
ment