6240B DC Voltage Current Source/Monitor Operation Manual
8.4 Calibration Points and Tolerance Ranges
8-5
Current HI limiter 30 A
0 A
+30.00 A5 nA
300 A +300.0 A50 nA
3 mA +3.000 mA 500 nA
30 mA +30.00 mA 5 A
300 mA +300.0 mA 50 A
1 A +1.000 A 500 A
4 A +1.000 A 1 mA
Current LO limiter 30 A
0 A
-30.00 A5 nA
300 A -300.0 A50 nA
3 mA -3.000 mA 500 nA
30 mA -30.00 mA 5 A
300 mA -300.0 mA 50 A
1 A -1.000 A 500 A
4 A -1.000 A 1 mA
Item Range
Calibration point
Tolerance range
Zero Full Scale