2-142 Chapter 2
System Verification and Performance Tests
Agilent 8753ES Performance Test Records
Agilent Technologies Company
Model 8753ES
Serial Number
Report Number
Option(s) Date
1. Test Port Output Frequency Range and Accuracy
Note: If your analyzer does not have Option 006, write “N/A” in all entries above 3 GHz.
CW Frequency
(MHz)
Min.
(MHz)
Results Measured
(MHz)
Max.
(MHz)
Measurement
Uncertainty
(MHz)
0.03 0.029 999 7 0.030 000 3 ± 0.000 000 050
0.3 0.299 997 0.300 003 ± 0.000 000 520
5.0 4.999 950 5.000 050 ± 0.000 007
16.0 15.999 840 16.000 160 ± 0.000 028
31.0 30.999 690 31.000 310 ± 0.000 054
60.999 999 60.999 389 61.000 609 ± 0.000 105
121.0 120.998 790 121.001 210 ± 0.000 207
180.0 179.998 200 180.001 800 ± 0.000 307
310.0 309.996 900 310.003 100 ± 0.000 528
700.0 699.993 000 700.007 000 ± 0.001 192
1 300.0 1 299.987 1 300.013 ± 0.002 212
2 000.0 1 999.980 2 000.020 ± 0.003 403 000
3 000.0 2 999.970 3 000.030 ± 0.005 104
4 000.0 3 999.960 4 000.040 ± 0.006 805
5 000.0 4 999.950 5 000.050 ± 0.008 506
6 000.0 5 999.940 6 000.060 ± 0.010 207