EasyManuals Logo

Agilent Technologies 8753ET Service Guide

Agilent Technologies 8753ET
617 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #221 background imageLoading...
Page #221 background image
Chapter 3 3-19
Adjustments and Correction Constants
Sampler Magnitude and Phase Correction Constants (Test 53)
8. Press . Make a
note of the insertion loss (found in the upper-right corner of the analyzer display).
Proceed to “Sampler Correction Constants Routine” on page 3-20.
Determine the Insertion Loss of the Cable at 1 GHz
(8753ET)
1. Press .
2. Press .
3. Refer to Figure 3-5.
Figure 3-5 First Connection for Insertion Loss Measurement (8753ET)
4. Perform a 1-port calibration by performing the following steps.
a. Press .
b. Press .
c. Connect the open (from the calibration kit) to the Reflection port as shown in Figure
3-5.
d. Press .
e. Press .
f. Connect the short to the Reflection port as shown in Figure 3-5.
g. Press .
h. Remove the short from the Reflection port, then connect the load to the Reflection
port as shown in Figure 3-5. Press , then .
Scale Ref
SCALE/DIV
.1
x1
Marker
MARKER 1
1 G/n
Preset Meas
REFLECTION
Center 1 G/n Span 50 M/µ
Cal
CALIBRATE MENU
REFLECTION 1-PORT
FORWARD: OPENS
OPEN (F)
DONE: OPENS
FORWARD: SHORTS
SHORT (F)
DONE: SHORTS
LOAD
DONE 1-PORT CAL

Table of Contents

Other manuals for Agilent Technologies 8753ET

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8753ET and is the answer not in the manual?

Agilent Technologies 8753ET Specifications

General IconGeneral
BrandAgilent Technologies
Model8753ET
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals