Chapter 3 3-35
Adjustments and Correction Constants
Correction Constants Retrieval Procedure
Correction Constants Retrieval Procedure
Required Equipment and Tools
By using the current EEPROM backup disk, you can download the correction constants
data into the instrument EEPROMs.
1. Insert the “EEPROM Backup Disk” into the analyzer’s disk drive.
2. Make sure the A9 switch is in the Alter position.
3. Press . Use the front panel knob to
highlight the file “N12345” where N12345 represents the file name of the EEPROM
data for the analyzer. On the factory shipped EEPROM backup disk, the filename is
“FILE1”.
4. Press to download the correction constants data into the
instrument EEPROMs.
5. Press and verify that good data was transferred to EEPROM by performing a
simple measurement.
6. Move the A9 switch back to the Normal position when you are done working with the
instrument.
Description HP/Agilent Part Number
EEPROM backup disk
Antistatic wrist strap 9300-1367
Antistatic wrist strap cord 9300-0980
Static-control table mat and earth ground wire 9300-0797
Save/Recall
SELECT DISK
INTERNAL DISK
RETURN
RECALL STATE
Preset

Table of Contents

Other manuals for Agilent Technologies 8753ET

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8753ET and is the answer not in the manual?

Agilent Technologies 8753ET Specifications

General IconGeneral
BrandAgilent Technologies
Model8753ET
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals