9-4 Chapter 9
Accessories Troubleshooting
Inspect the Accessories
Inspect the Accessories
Inspect the Test Port Connectors and Calibration Devices
1. Check for damage to the mating contacts of the test port center conductors and loose
connector bulkheads.
2. Inspect the calibration kit devices for bent or broken center conductors and other
physical damage. Refer to the calibration kit operating and service manual for
information on gaging and inspecting the device connectors.
If any calibration device is obviously damaged or out of mechanical tolerance, replace
the device.