Chapter 3 3-7
Adjustments and Correction Constants
Source Default Correction Constants (Test 44)
Source Default Correction Constants (Test 44)
Analyzer warm-up time: 30 minutes.
This internal adjustment routine writes default correction constants for the source power
accuracy.
1. Press
.
2. Observe the analyzer for the results of the adjustment routine:
If the analyzer displays *Source Def DONE, you have completed this procedure.
If the analyzer displays *Source Def FAIL, refer to Chapter 7 , “Source
Troubleshooting.”
Preset System
SERVICE MENU
TESTS
44 x1