11-4 Chapter 11
Error Terms
Error Correction
Error Correction
Use one of the following error-correction procedures depending on which analyzer you
have:
• 8753ES: “9. Uncorrected Port Performance” on page 2-50
• 8753ET: “10. Uncorrected Port Performance” on page 2-131
Error Term Inspection
Make a hardcopy of the measurement results:
1. Connect a printing or plotting peripheral to the analyzer.
2. Press and select the appropriate
peripheral to verify that the GPIB address is set correctly on the analyzer.
3. Press and then choose either or .
4. Press and title each data trace so that you can identify it
later.
For detailed information on creating hardcopies, refer to “Printing, Plotting, and Saving
Measurement Results” in your analyzer’s user’s guide.
If Error Terms Seem Worse than Typical Values
1. Perform a system verification to verify that the system still conforms to specifications.
2. If system verification fails, refer to Chapter 4 , “Start Troubleshooting Here.”