Chapter 11 11-5
Error Terms
Error Correction
Error Term Descriptions
The error term descriptions in this section include the following information:
significance of each error term
typical results following a full 2-port error-correction
guidelines to interpret each error term
The same description applies to both the forward (F) and reverse (R) terms. Reverse terms
do not apply to the 8753ET.
Table 11-1 Error Coefficient Arrays and Tests
Array Enhanced
Response
2-port
Test
Number
01 E
D
E
DF
32
02 E
S
E
SF
33
03 E
R
E
RF
34
04 E
X
E
XF
35
05
E
L
a
a. This term is used to generate the calibration
coefficients, but is not used during
measurement error correction.
E
LF
36
06 E
T
E
TF
37
07 E
DR
38
08 E
SR
39
09 E
RR
40
10 E
XR
41
11 E
LR
42
12 E
TR
43
Meaning of first subscript: Meaning of second subscript:
D: directivity F: forward
S: source match R: reverse
R: reflection tracking
X: crosstalk or isolation
L: load match
T: transmission tracking