EasyManua.ls Logo

Agilent Technologies 8753ET - Sampler Correction Constants Routine

Agilent Technologies 8753ET
617 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-20 Chapter 3
Adjustments and Correction Constants
Sampler Magnitude and Phase Correction Constants (Test 53)
5. Attach a type-N cable and a short to the Reflection port as shown in Figure 3-6.
Figure 3-6 Second Connection for Insertion Loss Measurement (8753ET)
6. Press .
7. Press .
8. Note the marker reading and divide it by 2. Record this calculated value as the
insertion loss of the cable. It will be used later in this procedure. Proceed to “Sampler
Correction Constants Routine,” next.
Sampler Correction Constants Routine
1. Press and
answer at the prompt.
2. The analyzer displays:
CONNECT <3 GHz SENSOR A TO PORT 1 (8753ES)
or
CONNECT <3 GHz SENSOR A TO REFL PORT (8753ET)
Make the connections as shown in Figure 3-7, using the 8482A power sensor.
Marker
MARKER 1
1
G/n
Sweep Setup
TRIGGER MENU
SINGLE
Preset System
TESTS
53 x1
EXECUTE TEST
YES

Table of Contents

Other manuals for Agilent Technologies 8753ET

Related product manuals