Chapter 1 1-3
Service Equipment and Analyzer Options
Required Tools and Equipment
Table 1-1 Agilent 8753ET Service Test Equipment
Required Equipment Critical Specifications HP/Agilent
Recommended Model
Use
a
Frequency Counter
Frequency: 300 kHz–3 GHz
(30 kHz–6 GHz for Option 006)
5350B/51B/52B P
Synthesized Sweeper
Maximum spurious input: <−30 dB
Residual FM: <20 kHz
83620A Option 001or
83712B Option 1E1
P
Spectrum Analyzer
Freq. Accuracy ±7 Hz 8563E A
Oscilloscope Bandwidth: 100 MHz
Accuracy: 10%
any T
Digital Voltmeter
Resolution: 10 mV any T
Tool Kit
No substitute 08753-60023 T
Power Meter (GPIB)
436A/437B/438A or
E4418A/E4419A
A, P, T
Power Sensor
Frequency: 300 kHz–3 GHz 8482A A, P, T
Power Sensor
(for Option 006)
Frequency: 3 GHz–6 GHz 8481A A, P, T
Photometer Tektronix J16 A
Photometer Probe
Tektronix J6503 A
Light Occluder Tektronix 016-0305-00 A
Printer HP ThinkJet, DeskJet,
LaserJet
P
Floppy Disk
one 3.5-inch formatted 1.44 MB
any
A
Calibration Kit Type-N,
50Ω
No substitute 85032B A, P
Calibration Kit 7 mm, 50Ω No substitute 85031B P
Verification Kit 7 mm
No substitute 85029B P
Low Pass Filter
>50 dB @ 2.96 Hz and passband
that includes 800 MHz
9135-0198 A
Fixed Attenuator (2)
Type-N 20 dB 8491A Opt. 020 P, T
Adapters (2)
APC-7 to Type-N (m) 11525A P, A
Adapter Type-N (m) to Type-N (m) 1250-1475 P
Adapter BNC to Alligator Clip 8120-1292 A
Adapter APC-3.5 (m) to Type-N (f) 1250-1750 A, P
Adapter APC-3.5 (f) to Type-N (f) 1250-1745 A, P